Background: Among the different factors affecting socket comfort, the pressure applied on residual limb tissues is a crucial parameter for the success or failure of any prosthetic device. However, only a few incomplete data are available on people with transfemoral amputation, in this regard. This work aims at filling this gap in the literature. Methods: Ten people with transfemoral amputation wearing 3 different socket designs were recruited in this study: 2 ischial containment sockets featured by proximal trim lines that contain the ischial tuberosity and ramus and greater trochanter, 2 subischial sockets with proximal trim lines under the ischium level, and 6 quadrilateral sockets with proximal trim lines that contain the greater trochanter and create a horizontal seat for the ischial tuberosity. The pressure values at the anterior, lateral, posterior, and medial areas of the socket interface were recorded during 5 locomotion tasks (ie, horizontal, ascent, and descent walking, upstairs and downstairs) by using an F-Socket System (Tekscan Inc., Boston, MA). Gait segmentation was performed by exploiting plantar pressure, which was acquired by an additional sensor under the foot. Mean and standard deviation of minimum and maximum values were calculated for each interface area, locomotion task, and socket design. The mean pressure patterns during different locomotion tasks were reported, as well. Results: Considering all subjects irrespective of socket design, the mean pressure range resulted 45.3 (posterior)-106.7 (posterior) kPa in horizontal walking; 48.3 (posterior)-113.8 (posterior) kPa in ascent walking; 50.8 (posterior)-105.7 (posterior) kPa in descent walking; 47.9 (posterior)-102.9 (lateral) kPa during upstairs; and 41.8 (posterior)-84.5 (anterior) kPa during downstairs. Qualitative differences in socket designs have been found. Conclusions: These data allow for a comprehensive analysis of pressures acting at the tissue-socket interface in people with transfemoral amputation, thus offering essential information for the design of novel solutions or to improve existing ones, in this field.

Quantitative analysis of interface pressures in transfemoral prosthetic sockets

Paterno, Linda
;
Truppa, Luigi;Ibrahimi, Michele;Rosini, Elisa;Ricotti, Leonardo;Menciassi, Arianna
2023-01-01

Abstract

Background: Among the different factors affecting socket comfort, the pressure applied on residual limb tissues is a crucial parameter for the success or failure of any prosthetic device. However, only a few incomplete data are available on people with transfemoral amputation, in this regard. This work aims at filling this gap in the literature. Methods: Ten people with transfemoral amputation wearing 3 different socket designs were recruited in this study: 2 ischial containment sockets featured by proximal trim lines that contain the ischial tuberosity and ramus and greater trochanter, 2 subischial sockets with proximal trim lines under the ischium level, and 6 quadrilateral sockets with proximal trim lines that contain the greater trochanter and create a horizontal seat for the ischial tuberosity. The pressure values at the anterior, lateral, posterior, and medial areas of the socket interface were recorded during 5 locomotion tasks (ie, horizontal, ascent, and descent walking, upstairs and downstairs) by using an F-Socket System (Tekscan Inc., Boston, MA). Gait segmentation was performed by exploiting plantar pressure, which was acquired by an additional sensor under the foot. Mean and standard deviation of minimum and maximum values were calculated for each interface area, locomotion task, and socket design. The mean pressure patterns during different locomotion tasks were reported, as well. Results: Considering all subjects irrespective of socket design, the mean pressure range resulted 45.3 (posterior)-106.7 (posterior) kPa in horizontal walking; 48.3 (posterior)-113.8 (posterior) kPa in ascent walking; 50.8 (posterior)-105.7 (posterior) kPa in descent walking; 47.9 (posterior)-102.9 (lateral) kPa during upstairs; and 41.8 (posterior)-84.5 (anterior) kPa during downstairs. Qualitative differences in socket designs have been found. Conclusions: These data allow for a comprehensive analysis of pressures acting at the tissue-socket interface in people with transfemoral amputation, thus offering essential information for the design of novel solutions or to improve existing ones, in this field.
2023
File in questo prodotto:
File Dimensione Formato  
Quantitative_analysis_of_interface_pressures_in.143.pdf

accesso aperto

Tipologia: PDF Editoriale
Licenza: Creative commons (selezionare)
Dimensione 1.16 MB
Formato Adobe PDF
1.16 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11382/556632
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
social impact